Click the Admin tab.
Click the Settings Tab.
Scroll to the HL7 Server Settings section.
The HL7 server settings define the server's HL7 configuration. To activate changes, edit the values and click Save. Restarting the server may cause current communications to abort.
The following HL7 parameters are configurable:
Setting |
Description |
HL7 Communications |
Enable/Disable HL7 communications. Disabling HL7 communications disables the listener only, but leaves the application running. A restart is required to make the new setting take effect. |
Port |
The TCP port number is the address of the HL7 process checks for TCP connections. |
HL7 Application ID |
Defined in the interface specification agreed to by both communicating HL7 entities. Refer to the HL7 Standard for format specifications. |
HL7 Facility ID |
Defined in the interface specification agreed to by both communicating HL7 entities. Refer to the HL7 Standard for format specifications. |
Send updates for non-Corrected studies |
Send ORM update messages to the HL7 device for studies that have not corrected to existing orders. |
Apply patient updates to studies newer than X days |
Apply ADT update messages to all studies and orders whose DICOM Study Date value is newer than X days. Set the value on the server connected to the HL7 device. The setting propagates to other servers. Set value to 0 to disable ADT updates. |
Supported HL7 messages and services, along with the mapping description files, are defined in the HL7 services table on the Devices page.